ISO/IEC 17025 reference laboratory
The company's specialised quality control and research and development laboratory, accredited to ISO/IEC 17025.

The company's laboratory is at the plant in the Charmshahr (Salarieh) Industrial Park, Varamin. It is responsible for quality control and assurance of raw materials and products, and for research, development and performance testing of the company's catalysts. Since January 2026 it has been accredited as a testing laboratory to ISO/IEC 17025:2017 by the National Accreditation Center of Iran (NACI), and it is a member of Iran's Strategic Technologies Laboratory Network.


Laboratory accreditation
ISO/IEC 17025:2017 Testing laboratory accreditation
- Issued by
- National Accreditation Center of Iran (NACI)
- No.
- NACI/Lab/2820
- Issued
- 20 January 2026
- Valid until
- 19 January 2029
Scope of accreditation
| Test | Samples | Applicable range | Reference |
|---|---|---|---|
| Determination of the specific surface area of solids by gas adsorption using the BET method | Porous solids and catalytic materials | 0.25 to 216 m²/g | ISO 9277:2022 |
| Particle size distribution of catalytic materials by laser light scattering (wet method) | Porous solids | 0.1 to 2000 µm | ASTM D4464:2020 |
| Nickel by atomic absorption spectrometry (flame) | Catalytic materials | From 3.3 % | LA-WI-010:1404 |
All three are test-oriented accreditations: the laboratory is accredited by the National Accreditation Center of Iran (NACI) to carry out these tests within these ranges. LA-WI-010:1404 is the laboratory's own documented test method.
Surface area and porositySurface area and porosity analyserMicromeritics · ASAP 2020Specific surface area, pore volume and pore size distribution, and mean pore diameter of mesoporous and microporous samples.
- Range
- Specific surface area from 0.25 to 216 m²/g
- Methods
- ISO 9277:2022
Surface area and porosityMercury porosimeterMicromeritics · AutoPore IVPore structure, surface, volume and size distribution.
- Range
- 0.003 to 1,100 µm
Particle sizeParticle size analyserFritsch · Analysette 22 NanoTec plusParticle size distribution of powders and suspensions.
- Range
- 0.1 to 2000 µm (wet method)
- Methods
- ASTM D4464:2020
Chemical analysisAtomic absorption spectrometerPerkinElmer · PinAAcle 900TElemental analysis of metals in many kinds of samples.
- Range
- At ppm and ppb levels; nickel from 3.3% within the accreditation scope (flame)
- Methods
- LA-WI-010:1404
Chemical analysisInfrared spectrometer (FTIR)Identification of functional groups in solid, liquid and gas samples.
Chemical analysisKarl Fischer titratorMettler-Toledo · C20Water content of solids, solvents, organic acids, alcohols, esters, ethers, anhydrides and hydrated inorganic salts.
Thermal analysisThermogravimetric analyser (TGA)Netzsch · TG 209 F1Mass change with temperature, phase changes and reaction enthalpy.
Physical propertiesViscometerBrookfield · DV-I PrimeViscosity of nanofluids, aqueous and hydrocarbon solutions, paints, resins such as epoxy, and adhesives.
Laboratory services for customers
The laboratory also offers specialised testing and assessment to industry and universities.
Laboratory services